The copper cycle (Experiment 1)

Five transformations that take copper metal all the way around and back — recover it, compute percent recovery.

Cu (s) copper metal Cu(NO₃)₂ (aq) blue solution Cu(OH)₂ (s) blue precipitate CuO (s) black solid CuSO₄ (aq) blue solution Cu (s) recovered! + HNO₃ + NaOH heat ~10 min + H₂SO₄ + Zn compare mass
Cu wire is accurately weighed (analytical balance) and sanded first. Zn is a large excess (2.00 g ±0.01 g, top-loading). Lost copper → recovery below 100%.

Glassware: TD vs TC & accuracy ladder

Pick the least precise tool that still meets the required accuracy. Precise = volumetric; rough = graduated / beaker.

less accurate more accurate Beaker ~±5% Grad. cylinder ~±1% Burette (TD) ±0.05 mL Vol. pipette / flask ±0.02 / ±0.08 mL TD = to DELIVER pipettes, Mohr pipettes, burettes TC = to CONTAIN volumetric flasks

Titration setup & reading the meniscus

Standardized titrant in the burette; analyte + phenolphthalein in the flask. Read the bottom of the meniscus at eye level.

burette titrant + phenolphthalein reading the meniscus read BOTTOM, eye level Rinse burette with DI then titrant · no bubble in the tip · ±0.02 mL Swirl · add dropwise near the endpoint · ≥3 concordant trials Endpoint = first faint pink that persists ~30 s

Ion-exchange column (Experiment 3)

Cu²⁺ displaces H⁺ (1 : 2) from the resin; the eluate H⁺ is titrated with NaOH to find [Cu²⁺].

reservoir DI rinse Cu²⁺ band resin (H⁺ form) glass wool (holds beads) eluate (H⁺) On the resin: Cu²⁺ + 2 R–H → R₂–Cu + 2 H⁺ Charge balance: 1 Cu²⁺ displaces 2 H⁺. Then titrate the eluate: mol Cu²⁺ = ½ × mol H⁺ = ½ × C(NaOH)·V(NaOH) [Cu²⁺] = mol Cu²⁺ ÷ 0.01000 L Error logic: sites not all H⁺-saturated → result too LOW leftover H⁺ not rinsed out → result too HIGH

Flame-test colours (Experiment 5)

Each metal ion emits a characteristic colour — a spectral fingerprint used to identify the cation.

Li⁺crimson-red
Na⁺intense yellow (589 nm)
K⁺lilac/violet (cobalt glass)
Ca²⁺orange-red
Sr²⁺scarlet-red
Ba²⁺pale green
Cu²⁺green / blue-green
Bgreen
E = hc/λ: shorter wavelength → higher energy (bluer). View K⁺ through cobalt-blue glass to filter out the overpowering Na⁺ yellow.

Accuracy vs precision

Accuracy = closeness to the true value. Precision = how tightly repeats cluster. They are independent.

accurate + precise precise, not accurate accurate, not precise neither
Precise but not accurate = a systematic error (e.g. a mis-read pipette). Scattered = random error.

Significant figures & melting-point purity

RuleExampleSig figs
Leading zeros — not significant0.00422
Captive zeros — significant1.0044
Trailing zeros after a decimal — significant2.503
×/÷ → fewest sig figs2.0000 × 10.03
+/− → fewest decimal places12.11 + 0.1→ 12.2
Melting point & purity: a pure solid melts over a narrow range near its literature value (acetaminophen ≈ 168–172 °C). Impurities depress and broaden the range.